The current state of fetal imaging by magnetic resonance imaging

Pavla Hanzlíková1, 2, 3, Dominik Vilímek1, 4, Radek Martinek4, Patricie Delongová5, 6, Jan Pavlíček7
1 Ústav radiodiagnostický, FN Ostrava
2 Ústav zobrazovacích metod, OU Ostrava
3 Radiologická klinika, Lékařská fakulta, Univerzita Palackého a FN Olomouc
4 Katedra kybernetiky a biomedicínského inženýrství, Vysoká škola báňská - Technická univerzita Ostrava
5 Ústav patologie, FN Ostrava
6 Ústav patologie, OU Ostrava
7 Klinika dětského lékařství, Lékařská fakulta, Univerzita Palackého a FN Olomouc

Magnetic resonance imaging (MRI), as a method of the second choice for fetal imaging, provides excellent spatial and contrast resolution for evaluating a wide range of pathological conditions, whether congenital or arising during pregnancy.

MR, as a method free of ionizing radiation, is together with ultrasound examination (UZ) a completely safe imaging method.

Because a standardized protocol scans the image, this imaging is not dependent on the examiner. It allows re-evaluation by the radiologist, clinical specialist and other doctors and specialists within the multidisciplinary team dedicated to the issue of prenatal fetal examination.

The benefit of MR imaging is a different principle of tissue morphological imaging compared to US examination. Another advantage of MR is the possibility of imaging not only morphologically but also using free water diffusion imaging - diffusion-weighting imaging (DWI), directional diffusion imaging (diffusion-tensor imaging -

DTI), metabolic composition imaging using spectroscopic methods (MR spectroscopy - MRS). Another option for viewing the fetus is a dynamic scan, where the vital functions of the fetus can be monitored over time, similar to an ultrasound. The use of contrast agents is a non-lege artis procedure in the Czech Republic and is not used in standard MR imaging of the fetus.

Keywords: fetus, magnetic resonance, fetal movement, DWI and DTI, MR section

Accepted: October 23, 2023; Published: December 11, 2023  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
Hanzlíková P, Vilímek D, Martinek R, Delongová P, Pavlíček J. The current state of fetal imaging by magnetic resonance imaging. Ces-slov Pediat. 2023;78(6):315-323. doi: 10.55095/CSPediatrie2023/052.
Download citation

References

  1. Benson CB, Doubilet PM. The history of imaging in obstetrics. Radiology 2014; 273(2 Suppl): S92-110. doi: 10.1148/radiol.14140238 Go to original source... Go to PubMed...
  2. Plunk MR, Chapman T. The fundamentals of fetal MR imaging: Part 1. Curr Probl Diagn Radiol 2014; 43(6): 331-346. doi: 10.1067/j.cpradiol.2014.05.014 Go to original source... Go to PubMed...
  3. Correa FF, Lara C, Bellver J, et al. Potential pitfalls in fetal neurosonography. Prenat Diagn 2006; 26(1): 52-56. doi: 10.1002/pd.1348 Go to original source... Go to PubMed...
  4. Knapp J, Tavares de Sousa M, Schönnagel BP. Fetal cardiovascular MRI - a systemic review of the literature: challenges, new technical developments, and perspectives. Rofo 2022; 194(8): 841-851. doi: 10.1055/a-1761-3500 Go to original source... Go to PubMed...
  5. McCarthy SM, Filly RA, Stark DD, et al. Magnetic resonance imaging of fetal anomalies in utero: early experience. AJR Am J Roentgenol 1985; 145(4): 677-682. doi: 10.2214/ajr.145.4.677 Go to original source... Go to PubMed...
  6. Tocchio S, Kline-Fath B, Kanal E, et al. MRI evaluation and safety in the developing brain. Semin Perinatol 2015; 39(2): 73-104. doi: 10.1053/j.semperi.2015.01.002 Go to original source... Go to PubMed...
  7. Meyers ML, Mirsky DM, Dannull KA, et al. Effects of maternal valium administration on fetal MRI motion artifact: a comparison study at high altitude. Fetal Diagn Ther 2017; 42(2): 124-129. doi: 10.1159/000450978 Go to original source... Go to PubMed...
  8. Malamateniou C, Malik SJ, Counsell SJ, et al. Motion-compensation techniques in neonatal and fetal MR imaging. AJNR Am J Neuroradiol 2013; 34(6): 1124-1136. doi: 10.3174/ajnr.A3128 Go to original source... Go to PubMed...
  9. De Wilde JP, Rivers AW, Price DL. A review of the current use of magnetic resonance imaging in pregnancy and safety implications for the fetus. Prog Biophys Mol Biol 2005; 87(2-3): 335-353. doi: 10.1016/j.pbiomolbio.2004.08.010 Go to original source... Go to PubMed...
  10. Shellock FG, Crues JV. MR procedures: biologic effects, safety, and patient care. Radiology 2004; 232(3): 635-652. doi: 10.1148/radiol.2323030830 Go to original source... Go to PubMed...
  11. Schenck JF. Safety of strong, static magnetic fields. J Magn Reson Imaging 2000; 12(1): 2-19. doi: 10.1002/1522-2586(200007)12 Go to original source...
  12. Mevissen M, Buntenkötter S, Löscher W. Effects of static and time-varying (50-Hz) magnetic fields on reproduction and fetal development in rats. Teratology 1994; 50(3): 229-237. doi: 10.1002/tera.1420500308 Go to original source... Go to PubMed...
  13. Wiskirchen J, Groenewaeller EF, Kehlbach R, et al. Long-term effects of repetitive exposure to a static magnetic field (1.5 T) on proliferation of human fetal lung fibroblasts. Magn Reson Med 1999; 41(3): 464-468. doi: 10.1002/(sici)1522-2594(199903)41 Go to original source...
  14. Levine D. Timing of MRI in pregnancy, repeat exams, access, and physician qualifications. Semin Perinatol 2013; 37(5): 340-344. doi: 10.1053/j.semperi.2013.06.011 Go to original source... Go to PubMed...
  15. Papaioannou G, Klein W, Cassart M, Garel C. Indications for magnetic resonance imaging of the fetal central nervous system: recommendations from the European Society of Paediatric Radiology Fetal Task Force. Pediatr Radiol 2021; 51(11): 2105-2114. doi: 10.1007/s00247-021-05104-w Go to original source... Go to PubMed...
  16. Colleran GC, Kyncl M, Garel C, Cassart M. Fetal magnetic resonance imaging at 3 Tesla - the European experience. Pediatr Radiol 2022; 52(5): 959-970. doi: 10.1007/s00247-021-05267-6 Go to original source... Go to PubMed...
  17. Macnaught G, Gray C, Walker J, et al. MRS: a potential biomarker of in utero placental function. NMR Biomed 2015; 28(10): 1275-1282. doi: 10.1002/nbm.3370 Go to original source... Go to PubMed...
  18. Coakley FV, Hricak H, Filly RA,et al. Complex fetal disorders: effect of MR imaging on management--preliminary clinical experience. Radiology 1999; 213(3): 691-696. doi: 10.1148/radiology.213.3.r99dc39691 Go to original source... Go to PubMed...
  19. Gatta G, Di Grezia G, Cuccurullo V, et al. MRI in pregnancy and precision medicine: a review from literature. J Pers Med 2021; 12(1). doi: 10.3390/jpm12010009 Go to original source... Go to PubMed...
  20. Simon EM, Goldstein RB, Coakley FV, et al. Fast MR imaging of fetal CNS anomalies in utero. AJNR Am J Neuroradiol 2000; 21(9): 1688-1698.
  21. Moradi B, Parooie F, Kazemi MA, et al. Fetal brain imaging: A comparison between fetal ultrasonography and intra uterine magnetic resonance imaging (a systematic review and meta-analysis). J Clin Ultrasound 2022; 50(4): 491-499. doi: 10.1002/jcu.23158 Go to original source... Go to PubMed...
  22. Kakish D, Tominna M, Krishnan A. Hemimegalencephaly: evolution from an atypical focal early appearance on fetal MRI to more conventional MR findings. Cureus 2022; 14(8): e27976. doi: 10.7759/cureus.27976 Go to original source... Go to PubMed...
  23. Vollbrecht TM, Luetkens JA. [Cardiac MRI of congenital heart disease : From fetus to adult]. Radiologie (Heidelb) 2022. doi: 10.1007/s00117-022-01062-y Go to original source... Go to PubMed...
  24. Brugger PC, Weber M, Prayer D. Magnetic resonance imaging of the normal fetal esophagus. Ultrasound Obstet Gynecol 2011; 38(5): 568-574. doi: 10.1002/uog.9002 Go to original source... Go to PubMed...
  25. Jaimes C, Yang E, Connaughton P, et al. Diagnostic equivalency of fast T2 and FLAIR sequences for pediatric brain MRI: a pilot study. Pediatr Radiol 2020; 50(4): 550-559. doi: 10.1007/s00247-019-04584-1 Go to original source... Go to PubMed...
  26. Masselli G, Vaccaro Notte MR, Zacharzewska-Gondek A, et al. Fetal MRI of CNS abnormalities. Clin Radiol 2020; 75(8): 640.e641-640.e611. doi: 10.1016/j.crad.2020.03.035 Go to original source... Go to PubMed...
  27. Corroenne R, Arthuis C, Kasprian G, et al. Diffusion tensor imaging in fetal brain: review to understand principles, potential and limitations of promising technique. Ultrasound Obstet Gynecol 2022. doi: 10.1002/uog.24935 Go to original source... Go to PubMed...
  28. Jiang S, Xue H, Counsell S, et al. Diffusion tensor imaging (DTI) of the brain in moving subjects: application to in-utero fetal and ex-utero studies. Magn Reson Med 2009; 62(3): 645-655. doi: 10.1002/mrm.22032 Go to original source... Go to PubMed...
  29. Chalouhi GE, Salomon LJ. BOLD-MRI to explore the oxygenation of fetal organs and of the placenta. BJOG 2014; 121(13): 1595. doi: 10.1111/1471-0528.12805 Go to original source... Go to PubMed...
  30. Cahill LS, Zhou YQ, Seed M, et al. Brain sparing in fetal mice: BOLD MRI and Doppler ultrasound show blood redistribution during hypoxia. J Cereb Blood Flow Metab 2014; 34(6): 1082-1088. doi: 10.1038/jcbfm.2014.62 Go to original source... Go to PubMed...
  31. Aertsen M, Diogo MC, Dymarkowski S, et al. Fetal MRI for dummies: what the fetal medicine specialist should know about acquisitions and sequences. Prenat Diagn 2020; 40(1): 6-17. doi: 10.1002/pd.5579 Go to original source... Go to PubMed...
  32. Chambers G, Shelmerdine SC, Aertsen M, et al. Current and future funding streams for paediatric postmortem imaging: European Society of Paediatric Radiology survey results. Pediatr Radiol 2022. doi: 10.1007/s00247-022-05485-6 Go to original source... Go to PubMed...

This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0), which permits use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.